DOTX-45 Dissolved Oxygen System Part 6 Configuration
DOTX-45 43 -
which the diagnostic system re-detects them. The
exception to this rule is the calibration failure. When a
calibration fails, no corrupt data is stored. Therefore, the
system continues to function normally on the data that was
present before the calibration was attempted.
After 30 minutes or if power to the transmitter is cycled, the
failure for calibration will be cleared until calibration is
attempted again. If the problem still exists, the calibration
failure will re-occur. Press ENTER to initiate view of the
highest priority failure. The display will automatically return
to normal after a few seconds.
PID Timer This function sets a timer to monitor the amount of time the
PID controller remains at 0% or 100%. This function only
appears if the PID controller is enabled. If the timer is set to
0000, the feature is effectively disabled. If the timer value is
set to any number other zero, a FAIL condition will occur if
the PID controller remains at 0% or 100% for the timer value.
If one of the relays is set to FAIL mode, this failure condition
can be signaled by a changing relay contact.
Press ENTER to initiate user entry mode, and the entire
value will flash. Use the UP arrow key to modify desired
value; range of value is 0-9999 seconds. Press ENTER to
store the new value.
Sim Out The Sim Out function allows the user to simulate the oxygen
level of the instrument in the user selected display range.
The user enters a ppm value directly onto the screen, and
the output responds as if it were actually receiving the signal
from the sensor. This allows the user to check the function
of attached monitoring equipment during set-up or
troubleshooting. Escaping this screen returns the unit to
normal operation. Press ENTER to initiate the user entry
mode, and the right-most digit of the value will flash. Use
arrow keys to modify desired value.
The starting display value will be the last read value of the
input. The output will be under control of the SIM screen
until the ESC key is pressed.
Note: If the HOLD function is engaged before the Sim Output
function is engaged, the simulated output will remain the
same even when the ESC key is pressed. Disengage the
HOLD function to return to normal output.