DOTX-45 Dissolved Oxygen System Part 6 – Configuration
DOTX-45 36 -
Note: If the user will not be changing solution conductivity
dramatically during the calibration process, leave a default
setting of 00.50 mS/cm. For ultrapure water applications, set
this value to 0.00. For sea water applications, set this value
to 53.00.
Iout#1 Mode This function sets analog output #1 to either track PPM or
mg/L oxygen (default), % Saturation, or enables the PID
controller to operate on the oxygen input in either PPM,
mg/L, or % Saturation. Press ENTER to initiate user entry
mode, and the entire value will flash. Use the UP arrow key
to modify the desired value; selections include 1- for oxygen
tracking or 2-PID for oxygen PID control. Press ENTER to
store the new value.
AC OPERATED UNITS ONLY
*Iout#2 Mode This function sets analog output #2 for either temperature
(default) or oxygen. Press ENTER to initiate user entry
mode, and the entire value will flash. Use the UP arrow key
to modify the desired value; selections include 1-C/F for
temperature, or 2-ppm for oxygen, or 3-% Saturation. Press
ENTER to store the new value.
AC OPERATED UNITS ONLY
*Rly A Mode Relay A can be used in three different ways: as a setpoint
control, as a fail alarm, or as a HI-LO alarm band. The three
settings for Rly A Mode are CON, FAIL and AL.
The CON setting enables normal control operation for Relay
A, with settings for setpoint, hysteresis, delay and phasing
appearing in the CONFIG menu automatically. See Figure
6-3 for further details.
The FAIL setting enables the fail alarm mode for Relay A.
Relay A will then trip on any condition that causes the FAIL
icon to be displayed on the LCD. Using this mode allows the
User to send alarm indications to other remote devices.
The AL setting allows two setpoints to be selected for the
same relay, producing a HI-LO alarm band. In this mode,
Relay A will trip inside or outside of the band, depending
upon the Phase selected. See Figure 6-4 for further details.